236 INSTRUMENTATION<br />

-<br />

Time<br />

Chloride<br />

min.<br />

Fig.7.21. Isotachophoretic separation of formate and acetate without (a) and with (b) a counter flow<br />

of electrolyte. The experiments were carried out in the operational system at pH 6 (Table 12.1) with<br />

glutamic acid as terminating electrolyte. Detection was carried out with a thermocouple mounted at<br />

a distance of about 50 cm from the injection point. The electric current was stabilized by the circuit<br />

shown in Fig.7.20. The thermocouple for regulation was mounted at a distance of about 25 cm<br />

from the injection point. The traces show that the electric current decreases (lower temperature) if<br />

a hot zone passes the regulating thermocouple. As soon as the terminating ion is below the regulation<br />

thermocouple, the electric current is stabilized at imin.. Trace (b) shows that equilibrium is achieved<br />

between the counter flow of electrolyte and the electric current. Traces (a) and (b) show that the<br />

isotachopherograms finally obtained are similar. These isotachopherograms are not given to show the<br />

usefullness of a counter flow of electrolyte, but only to demonstrate current stabilization via LDR<br />

and the possibility of using a counter flow of electrolyte (see Chapter 17).<br />

and the isotachopherogram that can be expected. The recording is performed with a<br />

thermometric detector (copper-constantan thermocouple). In this experiment, therefore,<br />

two thermocouples were mounted around the narrow-bore tube, one at the beginning<br />

of the narrow-bore tube, which was used for the LDR regulation during the counter flow<br />

of electrolyte period, and the other at the end of the narrow-bore tube, which was used<br />

for the detection. Because more advanced systems are considered later, further isotacho-<br />

pherograms with this type of regulation will not be shown. The experimental conditions<br />

for the isotachopherograms shown in Fig.7.21, however, will be given.<br />

The analysis was performed in the operational system at pH 6 (Table 12.1) with<br />

chloride as the leading ion and glutamate as the terminating ion. The aluminium block