26.12.2012 Views

Frans_M_Everaerts_Isotachophoresis_378342.pdf

Frans_M_Everaerts_Isotachophoresis_378342.pdf

Frans_M_Everaerts_Isotachophoresis_378342.pdf

SHOW MORE
SHOW LESS

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

THERMOMETRIC MEASUREMENTS 215<br />

vj c<br />

Kcl= c*L"<br />

(10.4)<br />

where Kcd is the calibration constant and L* (sec) is the zone length as detected between<br />

two successive signals obtained from the equipment if a zone boundary passes the detector. By<br />

means of a computer program (Chapter 4), the actual concentrations of the ionic species<br />

in the zone can be calculated in a well defined operational system. This means that once<br />

the calibration constant is known, the concentration of all ionic species in a sample can<br />

be calculated from the zone length. Not all calibration graphs for each ionic species have<br />

to be measured separately.<br />

In order to check the reproducibility and to determine simultaneously the calibration<br />

constant, K,,, quantitative experiments were carried out in different electrolyte systems<br />

with water as the solvent. The calibration constant is not a constant for all systems. Some<br />

factors, such as variations in the concentration of the leading electrolyte, temperature<br />

and changes in the current density, result in different potential gradients and hence affect<br />

the migration speed in the system. This effect produces different zone lengths for the<br />

same amounts of ionic species in different systems.<br />

10.3. THERMOMETRIC MEASUREMENTS<br />

10.3.1. Reproducibility<br />

In order to estimate the reproducibility, the zone length of formic acid (injected volume<br />

3 p1 of a 0.05 Nsolution) was measured ten times in different experiments. The leading<br />

electrolyte was 0.01 Nhistidine and 0.01 N histidine hydrochloride. The pH of the<br />

solution was thus 6.02. The current was stabilized at 70 FA. The terminator was 0.005N<br />

glutamic acid (adjusted to pH 6 by addition of Tris).<br />

The average zone length found was I* = 31 1 sec from ten experiments and the<br />

average deviation was 4 sec. Owing to the asymmetry of the step response, the zone length<br />

depends on the terminator used. Some experiments were therefore carried out with<br />

the same sample but with a different terminator (acetic acid). The average zone length<br />

then found was i* = 307 sec from five experiments and the average deviation was<br />

3 sec. No significant differences were found when the values found in the experiments<br />

with glutamic acid and acetic acid were compared. Glutamic acid was therefore used<br />

as the terminating ion in the other experiments.<br />

Later experiments showed that one has to be careful if glutamic acid is used as a<br />

terminating ion because it is a strong acid and hence mixed zones can easily be formed,<br />

which cannot be separated further.<br />

10.3.2. Calibration constant<br />

The calibration constant was determined from experiments carried out with histidine<br />

(0.01 N)/histidine hydrochloride (0.01 N) (pH 6.02) as the leading electrolyte. The<br />

current was stabilized at 70pA. All zone lengths are given in Table 10.1. The third

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!