26.12.2012 Views

Frans_M_Everaerts_Isotachophoresis_378342.pdf

Frans_M_Everaerts_Isotachophoresis_378342.pdf

Frans_M_Everaerts_Isotachophoresis_378342.pdf

SHOW MORE
SHOW LESS

Create successful ePaper yourself

Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.

316 COUNTER FLOW OF ELECTROLYTE<br />

detector is not needed. This regulation, however, places demands on the electrolytes of<br />

the operational system, which must be of high purity and also of carefully selected<br />

concentration. If the terminating electrolyte is impure, an impurity may influence the<br />

final recording if its effective mobility is greater than that of the terminating ion (see<br />

eqn. 7.4.). Both the pH and the concentration of the terminating electrolyte may have<br />

an influence on the regulation of the counter flow of electrolyte. At the position<br />

where the terminator is present in the narrow bore (1 mm) of the injection system (see<br />

Fig.7.5.), a well defined potential gradient is obtained as a result of the electrophoretic<br />

driving current chosen and the concentration of the electrolyte present (Ohm’s law). In<br />

isotachophoretic experiments without a counter flow of electrolyte, the concentration<br />

of the terminator has hardly an influence, assuming that it is present in such a concentra-<br />

tion that it is able to adapt to the isotachophoretic concentration, which is determined<br />

by the concentration of the leading electrolyte chosen (ratio of the anion and cation<br />

concentrations). If a counter flow of electrolyte is now applied that is regulated by the<br />

total potential gradient between the electrodes, the total resistance over the bore (1 mm)<br />

filled with terminating electrolyte changes, owing to the counter flow of electrolyte applied,<br />

if the composition of the terminating electrolyte is not already that which it would<br />

finally attain as a result of the isotachophoretic conditions.<br />

If the concentration of the terminating electrolyte is too low, then in a long run the<br />

potential gradient will decrease, because the final conductivity* is higher. This results in a<br />

movement of the sample zones in the direction of the detector. If the Concentration of<br />

the terminating electrolyte is too high, then the sample zones will be pushed back,<br />

although this effect is much smaller and generally has a less important effect on the final<br />

results.<br />

If a later experiment with a counter flow of electrolyte is to be carried out, it is<br />

preferable to rinse the reservoir fdled with terminating electrolyte and to re-fill it with<br />

a fresh solution, although this is not always necessary in experiments without a counter<br />

flow of electrolyte**. Moreover, it was found experimentally to be preferable to fill the<br />

reservoir at least three-quarters full with the terminating electrolyte, otherwise impurities<br />

can penetrate into the narrow-bore tube more easily. If impurities are present in the<br />

leading electrolyte, they may influence the final recording (see eqn. 7.3).<br />

Particular attention should be paid to the electrolyte present in the counter electrode<br />

compartment at the side of the semi-permeable membrane where the electrode is<br />

mounted. In experiments without a counter flow of electrolyte, double-distilled water is<br />

suitable because the membrane is not polluted by the counter ion chosen. Especially if<br />

the equipment is used for various operational systems, the mutual effect of the various<br />

counter ions is minimal. If experiments with a counter flow of electrolyte are considered,<br />

the pH shift due to the membrane and the electrode reaction may influence the final<br />

result more quickly. If double-distilled water is still preferred, this compartment must<br />

be rinsed continuously during the analysis. If leading electrolyte is present, this compart-<br />

ment must still be rinsed and re-filled after each experiment, assuming that the time of<br />

analysis is not excessive, otherwise it must also be supplied continuously with fresh<br />

electrolyte.<br />

*The buffer ions are flushed into the canals filled with terminating ions.<br />

**If experiments are carried out at a low concentration (0.01 M), it is preferable to start each experiment<br />

again with a fresh solution of the terminating electrolyte.

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!