26.12.2012 Views

Frans_M_Everaerts_Isotachophoresis_378342.pdf

Frans_M_Everaerts_Isotachophoresis_378342.pdf

Frans_M_Everaerts_Isotachophoresis_378342.pdf

SHOW MORE
SHOW LESS

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

EQUIPMENT 229<br />

mounted equiplanar, we found it advantageous to have this capacitor between the<br />

measuring electrodes; possibly an exact equiplanar construction is not always possible.<br />

Owing to the high potentials applied, some leak current will decrease the resolution<br />

after a small series of experiments. The signals derived from the transformer are handled<br />

electronically and result in a trace on the potentiometric recorder. This trace has a<br />

continuous stepwise character if the isotachophoretic zones pass the detector. If the<br />

trace does not have a continuous stepwise character, e.g, if a drift is obtained or dips<br />

and/or overshoots are recorded, something is wrong. A drift of the base line is obtained<br />

if, for instance, impurities are present that are more mobile than the leading ion, the<br />

buffer capacity of the counter ion is not sufficient or some electrode reaction occurs at<br />

the micro-sensing electrodes. Dips (or negative steps) can be expected if the buffer<br />

capacity of the counter ion is not sufficient, if an enforced isotachophoretic system is<br />

obtained or if the electrodes are coated with a polymer as a result of an electrode<br />

reaction or the physical adsorption of any material. Overshoots can be expected if the<br />

buffer capacity of the counter ion is not sufficient, if the temperatures of two adjacent<br />

zones are too different or if an electrode reaction occurs.<br />

A modified Brandenburg (Thornton Heath, Great Britain) power supply of the<br />

alpha-series is used. It is modified in such a way that it not only can be applied as a<br />

constant-voltage source (+30 kV), but for the isotachophoretic experiments can also be<br />

applied as a current-stabilized power supply (lt30 kV). Various current-stabilized power<br />

supplies, however, are commercially available nowadays, even up to 60 kV.<br />

Between the injection block, shown in detail in Figs.7.5 and 7.6, a six-way tap<br />

(Figs.7.2-7.4), whch can easily be removed without changing the narrow-bore tube if<br />

necessary, is mounted. In the equipment shown, an injection can also be made with a<br />

normal commercially available micro-syringe, as the sample can be introduced via the<br />

tap. Thus this instrument combines all the advantages of taps and syringes. Also, instead<br />

of the cylindrical counter electrode compartment (Figs.7.8 and 7.1 5), a counter electrode<br />

compartment with a flat membrane (Figs.7.9 and 7.10) can be used. These electrode<br />

compartments can easily be changed if necessary without any problems or the need to<br />

fit a new narrow-bore tube.<br />

All components of the isotachophoretic equipment shown in Fig.7.15 are replaceable<br />

because no adhesive is applied, rubber O-rings and screw-threads being used for clamping.<br />

We found that it is sometimes necessary to replace the narrow-bore tube as their life was<br />

found to vary from several years to only a few months. A narrow-bore tube needs to be<br />

changed if a decreased resolution cannot be improved.<br />

Owing to differences in the behaviour of the various operational systems, or compounds<br />

of the sample, even the ‘inert’ PTFE can become coated with material that is not easy<br />

to remove, and the resolution may decrease. Impurities, possibly building up over a long<br />

period, that are adsorbed on the walls of the injection system or the counter electrode<br />

compartment have less influence on the detection or the separation itself, because the<br />

bores in these parts of the instruments are considerably greater. The conductivity probe,<br />

if washing with a non-ionic detergent gives no improvement, can easily be cleaned with<br />

some metal polish and a cotton thread.

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!