26.12.2012 Views

Frans_M_Everaerts_Isotachophoresis_378342.pdf

Frans_M_Everaerts_Isotachophoresis_378342.pdf

Frans_M_Everaerts_Isotachophoresis_378342.pdf

SHOW MORE
SHOW LESS

Create successful ePaper yourself

Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.

INTRODUCTION 25 1<br />

the various ionic species in the operational systems and they indicate which ions can be<br />

separated in a given length of narrow-bore tube, assuming that the concentration differences<br />

are not too great. If the concentration differences are too great, a longer narrowbore<br />

tube or a counter flow of electrolyte must be used (see Chapter 17). It should be<br />

noted that no corrections for the temperatures of consecutive zones have been made to<br />

the results presented from either the thermometric or conductivity detector.<br />

If the influence of the counter ion chosen is great (high effective mobility in the opera-<br />

tional system chosen), greater differences in effective mobility between the sample ions<br />

are needed for a complete separation. If the pH of the consecutive zones increases<br />

regularly (in anionic separations) or decreases regularly (in cationic separations), small<br />

differences in effective mobility are often sufficient because once the ions diffuse into<br />

the zone where the pH is higher (anionic separations) or lower (cationic separations), they<br />

may attain a greater effective mobility owing to dissociation. More attention is paid to<br />

this phenomenon in Chapter 9.<br />

It is sometimes easier and quicker to apply two or more operational systems, as the<br />

equipment can be rinsed in a few minutes and is then ready for another operational<br />

system, than to try to carry out a complete separation in one operational system. More<br />

attention is paid to this aspect in Chapter 11. Sometimes only the concentration of the<br />

operational system needs to be changed. The sulphate ion, for instance, in the operational<br />

system at pH 6 (Table 12.1) moves behind the chloride ion (concentration of the leading<br />

ion, C1- = 0.01 N), while it has a greater mobility than the chloride ion if the concentra-<br />

tion of the ‘leading’ chloride ion is changed to 0.001 N. Another example is given in<br />

Fig.12.7. In principle, we do not recommend the use of too long a narrow-bore tube,<br />

because the voltages needed are too high and electroendosmosis may dominate the<br />

separation.<br />

If two types of detectors are available, in many instances the analysis can be carried<br />

out, in spite of stable mixed zones (see Fig.6.33), in one operational system. From time<br />

to time the equipment used in isotachophoretic analyses must be washed well with a non-<br />

ionic detergent followed by thorough rinsing with double-distilled water. Adsorption of<br />

many types of compounds may influence the detection (see Chapter 6), because amongst<br />

other effects the {-potential may be changed. Because the resolution of both the W<br />

absorption detector and the conductivity detector (ax. method) decreases in such<br />

instances, an electrode reaction only must be rejected. We prefer the a.c. method to the<br />

d.c. method because the detector indicates more quickly if something is going wrong and<br />

measures can be taken directly (see Chapter 6). We recommend that, before a series of<br />

analyses, a test mixture of anions or cations should be examined, because it can be seen<br />

very quickly if the resolution and reproducibility are adequate. If the test mixture indi-<br />

cates that non-reproducible data can be expected, appropriate measures must be taken,<br />

as shown in Fig.8.1.<br />

In our analyses, we pay special attention to the pH and the concentration of the<br />

terminating electrolyte, although from various papers one may obtain the impression that<br />

this is unnecessary. If too high a concentration of the terminating electrolyte is chosen,<br />

eluting effects due to the impurities can be expected, the sample zones may still migrate<br />

if a 100% counter flow of electrolyte is present (while the regulation is made via the

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!