4 <strong>Update</strong>s via Instant Pilot (G4208A)<br />
Updating the <strong>Firmware</strong> Using The Wizard<br />
Updating the <strong>Firmware</strong> Using The Wizard<br />
NOTE<br />
Only one firmware revision (set) should be stored on the USB Flash Drive. If more than one<br />
firmware version for a module is available, the Instant Pilot will always take the most recent<br />
firmware version. In this case, delete those versions not required.<br />
CAUTION<br />
Do not disconnect the Instant Pilot or the USB Flash Drive while a firmware update is<br />
running. Otherwise the module may become unusable.<br />
To open the Maintenance - System information, press More from the Welcome<br />
screen and select Maintenance.<br />
<strong>Update</strong> Wizard allows<br />
selective updates<br />
USB Flash Drive must be<br />
active<br />
Figure 60<br />
<strong>Firmware</strong> <strong>Update</strong> Screen - <strong>Update</strong> Wizard<br />
The <strong>Update</strong> Wizard allows you to define the modules to be updated. This screen<br />
also shows the installed firmware revision versus the available.<br />
1 Press <strong>Update</strong> Wizard and then Cont., or Abort to cancel the update process.<br />
2 The next screen (Figure 61) shows all modules, their installed firmware<br />
revisions and the available firmware revisions on the USB Flash Drive.<br />