Firmware Update - Agilent Technologies
Firmware Update - Agilent Technologies
Firmware Update - Agilent Technologies
You also want an ePaper? Increase the reach of your titles
YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.
<strong>Update</strong>s via Instant Pilot (G4208A) 4<br />
<strong>Update</strong> Information for A.05.1x <strong>Firmware</strong><br />
NOTE<br />
After the update/activation of the resident firmware<br />
• the Instant Pilot update screen shows an error (timeout) and/or<br />
• a screen may show up for some seconds showing "Waiting for module". Ignore this message.<br />
• the module's LED is blinking (resident mode)<br />
• this module disappears from the system (no communication with A.06.xx)<br />
4 Press 'Abort' and re-enter the firmware update screen.<br />
5 Select the next LC module.<br />
6 Repeat steps 3) to 5) until all LC modules are done.<br />
7 <strong>Update</strong> the Instant Pilot to A.05.1x (has only main firmware).<br />
8 Press Exit and restart the Instant Pilot.<br />
After re-boot, the Instant Pilot shows all resident LC modules.<br />
9 Enter the firmware update screen (More - Maintenance).<br />
10 Select the first resident LC module (not the Controller!).<br />
11 Press Single and start the main firmware update of this module.<br />
12 When finshed press Exit and No to prevent a re-boot.<br />
13 Select the next resident LC module and repeat steps 11) to 12) until all LC<br />
modules are done.<br />
After the update a screen may show up for some seconds showing "Waiting<br />
for module". Ignore this message. It will be closed after some seconds.<br />
14 Press Exit and restart the Instant Pilot.<br />
Upgrade to Revision from A.05.xx to B.02.xx<br />
When the Instant Pilot needs to be upgraded to a B.01.xx firmware to be<br />
compatible with <strong>Agilent</strong> 1100/1200 series modules working on<br />
A.06.xx/B.01.xx, follow the procedue below.<br />
1 Enter the firmware update screen (More - Maintenance).<br />
2 Select the first LC module (not the Controller!).<br />
3 Press Single and start the resident firmware update of this module.<br />
1100/1200/1290 Series <strong>Firmware</strong> <strong>Update</strong> Guide 105