06.11.2014 Views

Firmware Update - Agilent Technologies

Firmware Update - Agilent Technologies

Firmware Update - Agilent Technologies

SHOW MORE
SHOW LESS

Create successful ePaper yourself

Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.

<strong>Update</strong>s via LAN/RS-232 <strong>Firmware</strong> <strong>Update</strong> Tool 2<br />

Solving Problems<br />

Multiple System <strong>Update</strong>s Not Possible<br />

With a single user licence or using the Instrument Utilities, only a single<br />

system can be updated.<br />

Instant Pilot Screen Is Dark After <strong>Firmware</strong> <strong>Update</strong><br />

If the Instant Pilot’s screen stays dark after a firmware update while the LED<br />

is green, the Instant Pilot<br />

• has started its screen saver function - press a key to re-activate the display.<br />

or<br />

• has booted in resident mode. In this mode just a basic system is loaded<br />

without the graphics drivers. Start the LAN/RS-232 <strong>Firmware</strong> <strong>Update</strong> Tool<br />

2.7 and update the main firmware again. There is no other way to install<br />

the main firmware in this mode! See “Recovering” on page 78.<br />

Instant Pilot <strong>Update</strong> Does Not Finish<br />

When updating an Instant Pilot G4208A that is included in a system with<br />

G1315C DAD/G1365C MWD and G1316 TCC, the update may fail during<br />

update of resident or main firmware. If this happens,<br />

• turn off the G1316 TCC<br />

• perform the update of the Instant Pilot G4208A<br />

• turn on the G1316 TCC.<br />

This problem is caused by the initial A.06.02 firmware of the G1316 TCC and<br />

will be solved in the next firmware release of the G1316 TCC.<br />

Firewall or Anti-Virus Programs may Block the Tool<br />

If you get connect problems to the modules, please check your Firewall or<br />

Anti-Virus program setup. It might block the LAN/RS-232 <strong>Firmware</strong> <strong>Update</strong><br />

Tool 2.7.<br />

1100/1200/1290 Series <strong>Firmware</strong> <strong>Update</strong> Guide 83