<strong>Update</strong>s via Control Module (G1323A/B) 5<br />
Updating the HPLC Module <strong>Firmware</strong> via Control Module<br />
Updating the HPLC Module <strong>Firmware</strong> via Control Module<br />
NOTE<br />
If you want to upgrade the firmware of modules to revision A.06.xx refer to “<strong>Firmware</strong><br />
<strong>Update</strong> to A.06.xx” on page 63 prior to start the process.<br />
NOTE<br />
CAN slaves (CAN-valves, UIB or automation interface) have only main firmware.<br />
1 Copy the DLB files (HPLC module firmware) to your PC card’s root<br />
directory.<br />
2 Insert the PC card into the Control Module and restart the Control Module<br />
by pressing m-key and Restart on Analysis screen.<br />
3 Press Views-System-Records and select the module you want to update.<br />
1100/1200/1290 Series <strong>Firmware</strong> <strong>Update</strong> Guide 119