Introduction 1<br />
<strong>Firmware</strong> Description<br />
NOTE<br />
<strong>Update</strong> of main system can be done in the resident system only (except the Control Module<br />
G1323A/B).<br />
<strong>Update</strong> of the resident system can be done in the main system only.<br />
File Naming Convention of <strong>Firmware</strong><br />
File naming convention of firmware<br />
Module Type of firmware Name<br />
Pumps, Autosampler, Detectors,<br />
Column Compartment<br />
Pumps, Autosampler, Detectors,<br />
Column Compartment<br />
Main<br />
Resident<br />
xxxxA_Avvv_bbb.DLB<br />
xxxxA_Bvvv_bbb.DLB<br />
res_Avvv_bbb.DLB<br />
res_Bvvv_bbb.DLB<br />
G4208A Instant Pilot Main G4208A_Bvvv_bbb.DLB<br />
G4208A Instant Pilot Resident RES_xxxxBvvv_bb.DLB<br />
G1323A Control Module Main LCAvvvEN.BIN<br />
G1323B Control Module Main LCBvvvEN.BIN<br />
where (e.g. 1315AB_A511_001.DLB)<br />
xxxxA is the product number, for example G1315A/B DAD<br />
A is the firmware revision<br />
vvv is the revision number, for example 511 is revision 5.11<br />
bbb is the build number, for example 001 is the first build<br />
NOTE<br />
All modules with on-board LAN (e.g. G1315/65C/D and G1314D/E) and the Instant Pilot<br />
require "B" firmware. This “B” firmware works with “A” firmware under specific<br />
requirements. See “Compatibility Matrix” on page 18 for details.<br />
1100/1200/1290 Series <strong>Firmware</strong> <strong>Update</strong> Guide 11