<strong>Update</strong>s via LAN/RS-232 <strong>Firmware</strong> <strong>Update</strong> Tool 2<br />

<strong>Firmware</strong> <strong>Update</strong> to A.06.xx<br />

Emulation Mode for 1200 modules with new product numbers<br />

The <strong>Agilent</strong> 1200 Series LC platform comprises several modules with new<br />

product numbers:<br />

• G1312B 1200 Series Binary Pump SL<br />

• G1314B 1200 Series Variable Wavelength Detector<br />

• G1314C 1200 Series Variable Wavelength Detector SL<br />

• G1316B 1200 Series Thermostatted Column Compartment SL<br />

• G1329B 1200 Series Autosampler SL<br />

• G1367B 1200 Series High Performance Autosampler<br />

• G1367C 1200 Series High Performance Autosampler SL<br />

• G1367D 1200 Series High Performance Autosampler SL+<br />

These LC modules cannot be controlled by older <strong>Agilent</strong> control software and<br />

third-party software. However all of these modules can be operated in the<br />

1100 Series Emulation Mode with older firmware revisions as described<br />

above. In addition the module has to be converted to a "TYPE", that will be<br />

recognized by the installed control software.<br />

For example 16-bit ChemStation revision A.10.02 and many third party<br />

software packages will not recognized the variable wavelength detector SL<br />

G1314C that offers a data acquisition rate of 55 Hz. To control this detector<br />

with 16-bit ChemStation revision A.10.02, it has to be converted to the<br />

G1314A using the LAN/RS-232 <strong>Firmware</strong> <strong>Update</strong> Tool 2.7 or above. To convert<br />

the module (see also “How to change the Type of a Module” on page 74):<br />

The ChemStation will recognize this detector as G1314A in the Emulation<br />

Mode, however it will operate with slower data acquisition rate. At any time<br />

this detector can be converted back to a G1314C and controlled with software,<br />

that support the data rate of 55 Hz and all new 1200 Series features.<br />

To convert the G1312B binary pump SL, new firmware sets for the G1312B<br />

have been created<br />

• 1312B_A601 for the A.06.01 release or<br />

• 1312B_A512 for the A.05.11/12 release or<br />

• 1312B_A510 for the A.05.08/10 release or<br />

• 1312B_A501 for the A.05.01/03 release<br />

1100/1200/1290 Series <strong>Firmware</strong> <strong>Update</strong> Guide 71