Introduction 1<br />

Procedures<br />

<strong>Update</strong> of firmware for MS hardware<br />

After upgrading ChemStation software (e.g. to revision A10.01) the software<br />

complains that the MS firmware needs a newer version.<br />

In this case open your \HPCHEM\MS\firmware\ folder and execute the<br />

MSUPDATE.EXE. This will open a telnet session and start the update of the<br />

MS.<br />

For this update the MS has to be connected to the ChemStation.<br />

1100/1200/1290 Series <strong>Firmware</strong> <strong>Update</strong> Guide 15