4 <strong>Update</strong>s via Instant Pilot (G4208A)<br />
<strong>Update</strong> Information for A.05.1x <strong>Firmware</strong><br />
<strong>Update</strong> Information for A.05.1x <strong>Firmware</strong><br />
NOTE<br />
Before you downgrade to revision A.05.1x, upgrade the Instant Pilot to the latest B.02.xx<br />
firmware revision. This will assure that the later upgrade from A.05.1x to B.02.xx works<br />
smoothly.<br />
NOTE<br />
The firmware revision A.05.1x for the Instant Pilot does not require an A.05.1x resident<br />
firmware.<br />
Using the Instant Pilot<br />
Prerequisites<br />
• the Instant Pilot should have firmware B.02.xx or newer installed (with<br />
latest resident firmware)<br />
• all target FW (main/resident) of all LC modules and the Instant Pilot must<br />
be on the USB Flash Drive (root)<br />
• no other firmware revisions must be on the USB Flash Drive (root)<br />
• all modules in the stack are connected via CAN<br />
• the Instant Pilot is plugged into one module of the stack<br />
• the USB Flash Drive is plugged into the Instant Pilot<br />
Downgrade from A.06.xx to A.05.11<br />
In those cases where a Instant Pilot has to be downgraded from B.xx.xx to<br />
A.05.1x to allow the operation with A.05.09/12 firmware installed on <strong>Agilent</strong><br />
1100/1200 series modules, follow the procedue below.<br />
1 Enter the firmware update screen (More - Maintenance).<br />
2 Select the first LC module (not the Controller!).<br />
3 Press 'Single' and start the resident firmware update of this module.<br />
104 1100/1200/1290 Series <strong>Firmware</strong> <strong>Update</strong> Guide