<strong>Update</strong>s via Instant Pilot (G4208A) 4<br />

<strong>Update</strong> Information for A.05.1x <strong>Firmware</strong><br />

Using the LAN/RS-232 <strong>Firmware</strong> <strong>Update</strong> Tool<br />

For detailes refer to “<strong>Update</strong>s via LAN/RS-232 <strong>Firmware</strong> <strong>Update</strong> Tool” on<br />

page 27.<br />

NOTE<br />

Use the LAN/RS-232 <strong>Firmware</strong> <strong>Update</strong> Tool 2.7 or above. Version 2.3 and below do not work<br />

with this procedure.<br />

NOTE<br />

If the Instant Pilot is in “Backward compatible mode ON”, it will not be seen by the<br />

LAN/RS-232 <strong>Firmware</strong> <strong>Update</strong> Tool. For updating firmware reconfigure the Instant Pilot to<br />

“Backward compatible mode OFF”.<br />

NOTE<br />

If your <strong>Agilent</strong> 1100/1200 system runs on firmware A.05.xx, the Instant Pilot is not<br />

recognized by the LAN/RS-232 <strong>Firmware</strong> <strong>Update</strong> Tool. In this case either one of the<br />

modules or the complete system must be upgraded to A.06.xx/B.01.xx.<br />

NOTE<br />

Before you downgrade the Instant Pilot to revision A.05.1x, upgrade the Instant Pilot to the<br />

latest B.xx.xx firmware revision. This will assure that the later upgrade from A.05.1x to<br />

B.xx.xx works smoothly.<br />

NOTE<br />

The prefered way of updating the Instant Pilot firmware should be the use of a USB Flash<br />

Drive. This is much faster than the use of the LAN/RS-232 <strong>Firmware</strong> <strong>Update</strong> Tool<br />

(4 minutes vs. 10 minutes via LAN vs. 70 minutes via RS-232).<br />

Use one of the following procedues:<br />

• “Downgrade from B.xx to A.05.1x” on page 108<br />

• “Upgrade from A.05.1x to B.xx” on page 108<br />

• “Upgrade from A.05.11 to A.05.1x (new build)” on page 109<br />

1100/1200/1290 Series <strong>Firmware</strong> <strong>Update</strong> Guide 107