7 <strong>Agilent</strong> 1100/1200 Modules - A Quick Overview<br />

3rd Party Control Software<br />

3rd Party Control Software<br />

If 3rd Party Control Software is used, these 3rd party vendors need to update<br />

their software packages to include the new 1200 modules. Until they provide<br />

patches, the new 1200 modules can be downgraded to 1100 functionality into<br />

the so called "1100 emulation mode" to specific firmware revisions. The new<br />

module's type will be changed to "A" via the new LAN/RS-232 <strong>Firmware</strong><br />

<strong>Update</strong> Tool 2.7 or above.<br />

Specific cases are<br />

• the G1329A ALS, that replaces the 1100 G1313A ALS. For systems that have<br />

a G1329A without a thermostat G1330B, 3rd party software may<br />

require/ask for a thermostat (temperature setting). For this case SN<br />

1200-010 provides a workaround by changing the type of the G1329A to<br />

G1313A via the LC Diagnostic Software (can be changed back at a later<br />

time).<br />

• The G1315C DAD SL/G1365C MWD SL and G1315D DAD/G1365D MWD<br />

cannot go below B.01.01/B.01.04 and will not work in 1100 emulation<br />

modes A.05.11/12 and below.<br />

• The G4208A Instant Pilot can be made invisible for 3rd party software using<br />

a switch in its configuration setup. It works only with firmware<br />

A.06.02/B.01.02.<br />

• The G4208A Instant Pilot can be downgraded to A.05.13 to work in 1100<br />

emulation modes A.05.11/12 and below.<br />