Firmware Update - Agilent Technologies
Firmware Update - Agilent Technologies
Firmware Update - Agilent Technologies
You also want an ePaper? Increase the reach of your titles
YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.
7 <strong>Agilent</strong> 1100/1200 Modules - A Quick Overview<br />
3rd Party Control Software<br />
3rd Party Control Software<br />
If 3rd Party Control Software is used, these 3rd party vendors need to update<br />
their software packages to include the new 1200 modules. Until they provide<br />
patches, the new 1200 modules can be downgraded to 1100 functionality into<br />
the so called "1100 emulation mode" to specific firmware revisions. The new<br />
module's type will be changed to "A" via the new LAN/RS-232 <strong>Firmware</strong><br />
<strong>Update</strong> Tool 2.7 or above.<br />
Specific cases are<br />
• the G1329A ALS, that replaces the 1100 G1313A ALS. For systems that have<br />
a G1329A without a thermostat G1330B, 3rd party software may<br />
require/ask for a thermostat (temperature setting). For this case SN<br />
1200-010 provides a workaround by changing the type of the G1329A to<br />
G1313A via the LC Diagnostic Software (can be changed back at a later<br />
time).<br />
• The G1315C DAD SL/G1365C MWD SL and G1315D DAD/G1365D MWD<br />
cannot go below B.01.01/B.01.04 and will not work in 1100 emulation<br />
modes A.05.11/12 and below.<br />
• The G4208A Instant Pilot can be made invisible for 3rd party software using<br />
a switch in its configuration setup. It works only with firmware<br />
A.06.02/B.01.02.<br />
• The G4208A Instant Pilot can be downgraded to A.05.13 to work in 1100<br />
emulation modes A.05.11/12 and below.<br />
156 1100/1200/1290 Series <strong>Firmware</strong> <strong>Update</strong> Guide