4 <strong>Update</strong>s via Instant Pilot (G4208A)<br />
Errors During <strong>Firmware</strong> <strong>Update</strong>s<br />
G1315C DAD / G1365C MWD Main <strong>Update</strong> Not Possible<br />
In a few cases an update of a stand-alone G1315C DAD / G1365C MWD via the<br />
Instant Pilot resulted into a problem when the module switches from<br />
main-mode to resident-mode (for main firmware update). The Instant Pilot did<br />
not receive power via the CAN bus. In this case<br />
• turn the DAD/MWD off, then on.<br />
• transfer the DAD/MWD into resident-mode.<br />
• start the update of the main firmware again<br />
• if update is still not possible,<br />
• turn the DAD/MWD off<br />
• add another 1100/1200 module (if available).<br />
• turn on the other module first, then the DAD/MWD<br />
• transfer the DAD/MWD into resident-mode.<br />
• start the update of the main firmware again<br />
• if update is still not possible, use “<strong>Update</strong>s via LAN/RS-232 <strong>Firmware</strong><br />
<strong>Update</strong> Tool” on page 27.<br />
The problem is related to the DAD/MWD main board and will be addressed by<br />
a new revision of the main board.<br />