4 <strong>Update</strong>s via Instant Pilot (G4208A)<br />

Errors During <strong>Firmware</strong> <strong>Update</strong>s<br />

G1315C DAD / G1365C MWD Main <strong>Update</strong> Not Possible<br />

In a few cases an update of a stand-alone G1315C DAD / G1365C MWD via the<br />

Instant Pilot resulted into a problem when the module switches from<br />

main-mode to resident-mode (for main firmware update). The Instant Pilot did<br />

not receive power via the CAN bus. In this case<br />

• turn the DAD/MWD off, then on.<br />

• transfer the DAD/MWD into resident-mode.<br />

• start the update of the main firmware again<br />

• if update is still not possible,<br />

• turn the DAD/MWD off<br />

• add another 1100/1200 module (if available).<br />

• turn on the other module first, then the DAD/MWD<br />

• transfer the DAD/MWD into resident-mode.<br />

• start the update of the main firmware again<br />

• if update is still not possible, use “<strong>Update</strong>s via LAN/RS-232 <strong>Firmware</strong><br />

<strong>Update</strong> Tool” on page 27.<br />

The problem is related to the DAD/MWD main board and will be addressed by<br />

a new revision of the main board.<br />

112 1100/1200/1290 Series <strong>Firmware</strong> <strong>Update</strong> Guide