2 <strong>Update</strong>s via LAN/RS-232 <strong>Firmware</strong> <strong>Update</strong> Tool<br />
<strong>Firmware</strong> <strong>Update</strong> to A.06.xx<br />
NOTE<br />
When downgrading to 1312B_A601 or 1312B_A512 or 1312B_A510 or 1312B_A501 the<br />
corresponding resident firmware (res_601 or res_A511 or res_A508 or res_A501) must be<br />
installed. Otherwise the interface protocols do not match.<br />
The Emulation Mode has been fully tested with <strong>Agilent</strong> software and based on<br />
this we would not expect issues with third-party software.<br />
Scenario A<br />
Situation: I have an 1100 Series LC system with binary pump, manual<br />
injector, thermostatted column compartment and variable wavelength<br />
detector (all firmware A.05.06) running with ChemStation A.10.01. Now I<br />
want to automate my system. Can I add a new 1200 Series standard<br />
autosampler? Do I have to upgrade the ChemStation Software?<br />
Solution: In general all 1200 Series Modules can be added to an existing 1100<br />
Series LC system. In this case there are two possible solutions:<br />
1 If the ChemStation can be upgraded to B.02.01 and the firmware to A.06.02,<br />
the system can be used including all features of the new 1200 Series<br />
autosampler (illumination).<br />
2 In the Emulation Mode the 1200 Series autosampler can be added to this<br />
system with ChemStation A.10.01. In this case its firmware must be<br />
downgraded to A.05.06. All modules should have the same firmware<br />
revision loaded. New features (e.g. Illumination of the sample<br />
compartment) are not available.<br />
Scenario B<br />
Situation: I have an 1100 Series LC system with isocratic pump, autosampler,<br />
thermostatted column compartment and variable wavelength detector (all<br />
firmware A.05.09) running with ChemStation A.10.02. I want to integrate a<br />
binary pump for gradient applications. Can I add the new 1200 Series binary<br />
pump SL G1312B now and upgrade the ChemStation software later?<br />
Solution: The binary pump SL will not be recognized by ChemStation Rev.<br />
A.10.02, but in the Emulation Mode this pump can be added to that system.<br />
72 1100/1200/1290 Series <strong>Firmware</strong> <strong>Update</strong> Guide