Downgrade of <strong>Agilent</strong> 1100/1200/1290 Modules 8<br />
Minimum firmware revision vs. mainboard revision<br />
1290 Infinity LC Modules<br />
Table 12 lists the various <strong>Agilent</strong> 1290 Infinity LC modules, their main board<br />
part numbers and the minimum firmware revisions that can be loaded.<br />
NOTE<br />
When downgrading or upgrading, the corresponding resident firmware should be loaded as<br />
well. For revisions refer to the <strong>Firmware</strong> Revision History document.<br />
Table 12<br />
Minimum firmware revision vs. mainboard revision (for notes see below table)<br />
Module Type Module description Mainboard<br />
part number<br />
Minimum<br />
firmware<br />
revision<br />
Notes<br />
Pump G4220A Infinity Binary Pump G4220-65800 B.06.23 1<br />
Detector G4212A Infinity Diode Array Detector G4212-65800 B.06.23 1<br />
Sampler G4226A Infinity Autosampler G4226-65000 A.06.14 1<br />
TCC G1316C Infinity Thermostatted Column Compartment G1316-65050 A.06.14<br />
(A.06.10)<br />
2<br />
Local<br />
Controller<br />
G4208A Instant Pilot with DE... serial numbers n/a B.02.09 3<br />
G4208A Instant Pilot with MY.... serial numbers n/a B.02.09 3<br />
Notes:<br />
1 Initial version, no further downgrade possible.<br />
2 Initial version for 1290 Infinity system. When used as G1316C in Method<br />
development system, minimum is A.06.10, see Table 11.<br />
3 Initial version, further downgrade possible for use with 1100/1200<br />
modules, see Table 11 on page 159.<br />
1100/1200/1290 Series <strong>Firmware</strong> <strong>Update</strong> Guide 163