<strong>Update</strong>s via Instant Pilot (G4208A) 4<br />
General Information<br />
Prerequisites<br />
• <strong>Agilent</strong> Instant G4208A connected to a <strong>Agilent</strong> 1100/1200/1290 Series<br />
series HPLC module<br />
• USB Flash Drive (<strong>Agilent</strong> part number G4208-68700).<br />
• firmware files from the <strong>Agilent</strong> web, see “<strong>Update</strong>s from the <strong>Agilent</strong> Web” on<br />
page 16.<br />
NOTE<br />
Since USB-memory sticks may vary from vendor to vendor or from type to type,<br />
incompatibilities can occur. In general, USB-memory sticks from Sandisk and Kingston<br />
should work. The USB-memory sticks must be FAT-16 formatted and without encryption.<br />
Use <strong>Agilent</strong> USB-Memory Stick Kit G4208-68700.<br />
NOTE<br />
In case the USB Flash Disk is not recognized by the Instant Pilot, use the operating system<br />
of your PC (Windows XP) to re-format the USB Flash Disk. Use as File System FAT (NOT<br />
FAT32).<br />
<strong>Update</strong> Sequence<br />
Using the <strong>Agilent</strong> Instant Pilot G4208A, the system is updated in the following<br />
order (the Instant Pilot takes care for this order):<br />
1 <strong>Update</strong> of the Instant Pilot<br />
The Instant Pilot can also be updated as described under “<strong>Update</strong>s via<br />
LAN/RS-232 <strong>Firmware</strong> <strong>Update</strong> Tool” on page 27.<br />
2 CAN slaves (CAN-valves, UIB or automation interface) (only main<br />
firmware)<br />
3 all resident firmware<br />
4 transfer of all modules into resident mode<br />
5 all module main firmware<br />
6 transfer of all modules into main mode.<br />