<strong>Update</strong>s via LAN/RS-232 <strong>Firmware</strong> <strong>Update</strong> Tool 2<br />

Recovering<br />

Recovering<br />

In case of power fail or LAN/RS-232 cable disconnected during the update to<br />

revision A.06xx, one or more modules will no longer visible in the system! It<br />

does not communicate with the modules that have A.06.xx firmware.<br />

Module has LAN or RS-232 communication<br />

If the module has LAN (slot for LAN interface) or RS-232 communication you<br />

have to<br />

1 Power off all modules.<br />

2 Add the LAN/RS-232 communication to the module that is no longer<br />

visible.<br />

3 Turn this module on.<br />

4 Start the firmware update to A.06.xx for this module.<br />

5 Power off the module.<br />

6 Re-install the LAN/RS-232 communication to the detector (DAD, MWD,<br />

FLD, VWD) or, if not available, to pump or autosampler (this order).<br />

7 Turn all modules on for normal operation.<br />

Module has no LAN communication<br />

If the module has no LAN communication (e.g. G1316A TCC) you have to<br />

1 Power off all modules except the module that hosts the LAN/RS-232<br />

communication.<br />

2 Downgrade this module to firmware A.05xx (resident and main).<br />

3 Turn the missing module on.<br />

4 Start the firmware update to A.06.xx for both modules.<br />

5 Turn all modules on for normal operation.<br />

1100/1200/1290 Series <strong>Firmware</strong> <strong>Update</strong> Guide 77