6 <strong>Update</strong>s via GPIB <strong>Firmware</strong> <strong>Update</strong> Tool<br />
General Information<br />
General Information<br />
This program should be used for<br />
• GPIB updates only<br />
NOTE<br />
This GPIB <strong>Firmware</strong> <strong>Update</strong> Tool is not compatible with <strong>Agilent</strong> 32-bit ChemStation B.xx.<br />
If you have to upgrade old GPIB system use these alternatives:<br />
• Do the upgrade of the firmware while a 16-bit ChemStation (A.10 or below) is installed<br />
(especially for G1600 CE instruments).<br />
• Use the LAN/RS-232 <strong>Firmware</strong> <strong>Update</strong> Tool 2.7 for upgrading 1100 modules via RS-232 or<br />
LAN (requires a LAN interface in one of the modules of a system) as described in<br />
“<strong>Update</strong>s via LAN/RS-232 <strong>Firmware</strong> <strong>Update</strong> Tool” on page 27.<br />
NOTE<br />
If you own a system with LAN or RS-232 communication, you should use “<strong>Update</strong>s via<br />
LAN/RS-232 <strong>Firmware</strong> <strong>Update</strong> Tool” on page 27.<br />
If you want to upgrade the firmware of modules to revision A.06.xx refer to “<strong>Firmware</strong><br />
<strong>Update</strong> to A.06.xx” on page 140 prior to start the process.<br />
The screens in this note may vary from your screens due to other firmware revision.<br />
The GPIB communication is no longer supported with <strong>Agilent</strong> ChemStation revision B.01.01<br />
and above.<br />
Prerequisites<br />
• GPIB <strong>Firmware</strong> <strong>Update</strong> Tool from the <strong>Agilent</strong> web<br />
• proper working ChemStation communication with the 1100<br />
system/modules<br />
• HPLC module specific firmware (main and resident) from the <strong>Agilent</strong> web<br />
134 1100/1200/1290 Series <strong>Firmware</strong> <strong>Update</strong> Guide