1 Introduction<br />
Compatibility Matrix<br />
Compatibility Matrix<br />
Figure 2 on page 21 and Figure 3 on page 22 show the timeline of<br />
introductions of <strong>Agilent</strong> 1100/1200 firmware versus the <strong>Agilent</strong> ChemStations<br />
since firmware revision A.05.01.<br />
NOTE<br />
<strong>Firmware</strong> used in a system should always be within a firmware revision set to assure full<br />
compatibility in the system. See also information on “Revision Numbering” on page 12<br />
Sets are:<br />
• A.05.01/A.05.03<br />
• A.05.04/A.05.05<br />
• A.05.06/A.05.09/A.05.10 (some modules remained on A.05.06)<br />
• A.05.11/A.05.12<br />
• A.06.01 and B.01.01<br />
• A.06.02/A.06.0x and B.01.02/B.01.0x<br />
• A.06.1x and B.06.1x/B.06.2x/B.06.3x<br />
• A.06.1x and B.06.2x (1290 Infinity LC System)<br />
• A.06.3x and B.06.3x (1290 Infinity LC System)<br />
NOTE<br />
If 1290 Infinity LC modules are added into existing 1100/1200 systems, then the 1100/1200<br />
must have A.06.1x and B.06.1x to work with A.06.1x and B.06.2x (1290). The same is true for<br />
adding 1100/1200 modules into a 1290 Infinity LC system.<br />
Table 2 on page 19 gives a quick overview.<br />
Figure 2 on page 21 shows the full compatibility with 1200 series modules.<br />
Figure 3 on page 22 shows the possible compatibility of 1200 series modules<br />
with 1100 series modules using the “1100/1200 Series Compatibility and<br />
Emulation Mode” on page 68.<br />
Table 3 on page 24 lists the requirements for the <strong>Agilent</strong> 1290 Infinity LC<br />
System.<br />
18 1100/1200/1290 Series <strong>Firmware</strong> <strong>Update</strong> Guide