Firmware Update - Agilent Technologies
Firmware Update - Agilent Technologies
Firmware Update - Agilent Technologies
Create successful ePaper yourself
Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.
<strong>Update</strong>s via Control Module (G1323A/B) 5<br />
PC-Card <strong>Update</strong> Mode<br />
Getting back into the PC-Card <strong>Update</strong> Mode<br />
To switch the Control Module back into this update mode proceed as follows:<br />
1 Insert the PC-card (LCB322EN.BIN or LCA210EN.BIN, CONFIG.SYS,<br />
UPDATE.EXE should be on the card).<br />
2 Enter the RECORDS screen (via Views-System) and select the G1323A/B.<br />
3 Press FW <strong>Update</strong> and then the m-key.<br />
4 Select Enable <strong>Update</strong> and leave the RECORDS screen.<br />
5 Re-boot by pressing the m-Key and selecting 'Restart'.<br />
1100/1200/1290 Series <strong>Firmware</strong> <strong>Update</strong> Guide 123