5 <strong>Update</strong>s via Control Module (G1323A/B)<br />

Changing the Serial Number of Control Module<br />

Changing the Serial Number of Control Module<br />

To change the serial number of an Control Module, for example when it shows<br />

serial number DE00000000 or CN00000000, proceed as follows:<br />

NOTE<br />

For entering CN serial numbers you need the UPDATE.EXE version dated March 16, 2001<br />

(version 3.19) added March 28, 2001 from the download servers. For details “Changing the<br />

Serial Number of Control Module” on page 124.<br />

1 Copy files LCB322EN.BIN, CONFIG.SYS, UPDATE.EXE to a PC-card.<br />

2 Insert the PC-card and re-boot. The display will show an update screen.<br />

Control Module <strong>Update</strong> Version 3.19 (c) 1996-2001 <strong>Agilent</strong> <strong>Technologies</strong><br />

1) program Control Module<br />

2) disable update at boot-up<br />

3) change serial number<br />

4) change display contrast<br />

6) display load address<br />

9) program 2nd rom<br />

0) exit<br />

select function: _<br />

3 Select (3) to change the serial number (preparation).<br />

4 Select (1) to start the programming.<br />

5 Select (n) LCB322EN.BIN (or newer, if available).<br />

Serial number invalid<br />

Press to input number found on rear side of Control Module<br />

_<br />

124 1100/1200/1290 Series <strong>Firmware</strong> <strong>Update</strong> Guide