2 <strong>Update</strong>s via LAN/RS-232 <strong>Firmware</strong> <strong>Update</strong> Tool<br />

Solving Problems<br />

Lost Connenction to Module<br />

If the update fails via LAN cross-over cable due to “lost connection to module”<br />

and the module remains in resident mode (flashing status LED), the process is<br />

interrupted.<br />

In this situation check your LAN setup, see “PC and <strong>Agilent</strong> ChemStation<br />

Setup” on page 40.<br />

Do not use a configuration like below. It uses the Default-IP address of the<br />

(192.168.254.11) of the module with the LAN-access. This setup does not work<br />

stable.<br />

Figure 47<br />

Problematic LAN Setup<br />

The RS-232 <strong>Update</strong> is So Slow<br />

Yes, the serial (RS-232) communication protocol is very slow compared to LAN<br />

communication. Therefore the LAN communication is the recommended<br />

communication protocol for firmware updates. See also “<strong>Update</strong><br />

Durations” on page 44.<br />

82 1100/1200/1290 Series <strong>Firmware</strong> <strong>Update</strong> Guide