4 <strong>Update</strong>s via Instant Pilot (G4208A)<br />
<strong>Update</strong> Information for A.05.1x <strong>Firmware</strong><br />
NOTE<br />
After the update/activation of the resident firmware<br />
• the Instant Pilot update screen shows list with the module(s) where<br />
• the update was interrupted or a timeout error<br />
• the module's LED is blinking (resident mode)<br />
• this module disappears from the system (no communication with A.05.xx)<br />
4 Press Abort.<br />
5 Enter the firmware update screen (More - Maintenance).<br />
6 Select the next LC module.<br />
7 Repeat steps 3) to 4) until all LC modules are done.<br />
8 <strong>Update</strong> the Instant Pilot to B.02.xx (main and resident firmware).<br />
9 Press Exit and restart the Instant Pilot.<br />
After re-boot, the Instant Pilot shows all resident LC modules.<br />
NOTE<br />
If the modules are not shown (continuoes “Scanning ...“), disconnect the Instant Pilot and<br />
reconnect.<br />
10 Enter the firmware update screen (More - Maintenance).<br />
11 Select the first LC module (not the Controller!).<br />
12 Press Single and start the main firmware update of this module.<br />
After the update a screen may show up for some seconds showing all LC<br />
modules where the update has been interrupted. Ignore this message. It will<br />
be closed after some seconds.<br />
13 When finshed press Exit and No to prevent a re-boot.<br />
14 Enter the firmware update screen (More - Maintenance).<br />
15 Select the next resident LC module and repeat steps 12) to 13) until all LC<br />
modules are done.<br />
After the update a screen may show up for some seconds showing all LC<br />
modules where the update has been interrupted. Ignore this message. It will<br />
be closed after some seconds.<br />
16 Press Exit and restart the Instant Pilot.<br />
106 1100/1200/1290 Series <strong>Firmware</strong> <strong>Update</strong> Guide