<strong>Update</strong>s via LAN/RS-232 <strong>Firmware</strong> <strong>Update</strong> Tool 2<br />

<strong>Firmware</strong> <strong>Update</strong> to A.06.xx<br />

<strong>Firmware</strong> Upgrade to A.06.02<br />

In March 2006, the <strong>Agilent</strong> 1200 series was introduced with firmware revision<br />

A.06.02. This revision is not compatible with all previous revisions. All<br />

modules in a system must be upgraded to A.06.02 prior to the connection to<br />

modules with A.06.02, or the A.06.02 modules must be downgraded to A.06.01<br />

or A.05xx prior to the connection to these modules.<br />

Table 6 on page 63 shows the compatibility impact. See also “Revision<br />

Numbering” on page 12 and “Compatibility Matrix” on page 18.<br />

The information below should be used when updating to firmware revision<br />

A.06.02 (from revision A.06.01 or A.05.xx and below).<br />

NOTE<br />

If you do not follow these instructions, you may loose the communication to a module. In<br />

this case follow the instruction “Recovering” on page 77.<br />

• Use only LAN/RS-232 <strong>Firmware</strong> <strong>Update</strong> Tool 2.7 or above to ensure correct<br />

order of update sequence, especially on systems with CAN slaves<br />

(CAN-valves, UIB or automation interface) included.<br />

• If a G1315C DAD or G1365C MWD (initial firmware B.01.02) is added to an<br />

existing system with firmware A.05.xx and below, first all modules with<br />

firmware A.05.xx and below must be upgraded to A.06.02 (the<br />

G1315C/G1365C not connected during the update).<br />

• Upgrade a G1323B Control Module to revision B.04.02 together with the<br />

modules that are updated to A.06.02. This will ensure that the Control<br />

Module recognizes old and new firmware modules. See also information on<br />

“<strong>Update</strong> Sequence” on page 29, “<strong>Update</strong> Durations” on page 44 and<br />

Figure 26 on page 49.<br />

NOTE<br />

If those modules that need to be upgraded to A.06.xx do not have a LAN interface, then a<br />

• LAN interface (see “<strong>Firmware</strong> <strong>Update</strong> via LAN” on page 30)or<br />

• serial cable (see “RS-232 Cable requirement” on page 31) or<br />

• a Control Module G1323B<br />

is required.<br />

1100/1200/1290 Series <strong>Firmware</strong> <strong>Update</strong> Guide 67