6 <strong>Update</strong>s via GPIB <strong>Firmware</strong> <strong>Update</strong> Tool<br />

Updating Module <strong>Firmware</strong><br />

Updating Module <strong>Firmware</strong><br />

• A properly installed LC ChemStation is required<br />

• Download the GPIB <strong>Firmware</strong> <strong>Update</strong> Tool and the module firmware files<br />

to your PC as described in “<strong>Update</strong>s from the <strong>Agilent</strong> Web” on page 16.<br />

1 Check that the 1100 System Access is configured using the Configuration<br />

Editor (Windows Start - <strong>Agilent</strong> ChemStation - Configuration Editor):<br />

• Configure instruments<br />

• Instrument Type = Modular (3D) LC System<br />

• Dialog box: Instrument x - Device Configuration<br />

• 1100 System Access with correct GPIB address of the module that is<br />

connected via GPIB (typically a detector) must be set<br />

• GPIB must be chosen (depends on your communication path)<br />

2 Start FWUPGRAD.EXE from the directory \FWUPGRAD.<br />

3 For all modules identified under a certain instrument number the program<br />

displays an icon, the product number of the instrument, the serial number<br />

and main system firmware revisions.<br />

Control Module<br />

version B only!<br />

136 1100/1200/1290 Series <strong>Firmware</strong> <strong>Update</strong> Guide