5 <strong>Update</strong>s via Control Module (G1323A/B)<br />

Changing the Serial Number of Control Module<br />

Message ‘Serial Number Invalid’<br />

To avoid a "Serial Number Invalid" message during update of Control Modules<br />

that have serial number starting with CN... and software revision B.02.02<br />

installed, you must copy the latest UPDATE.EXE (revision 3.19 or above, dated<br />

03/16/01 or later) onto the same flash card with the B.03.22 or later firmware.<br />

All updates from this card will then be processed the normal way without<br />

showing the described error message.<br />

If the error already occurred, press and type the serial number<br />

printed on the rear side of the Control Module into the shown mask. Then<br />

press and to acknowledge the serial number.<br />

The message should not occur again, after an update to revision B.03.01 or<br />

above.<br />

NOTE<br />

The described problem does not occur during remote updates via PC.<br />