Firmware Update - Agilent Technologies
Firmware Update - Agilent Technologies
Firmware Update - Agilent Technologies
You also want an ePaper? Increase the reach of your titles
YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.
2 <strong>Update</strong>s via LAN/RS-232 <strong>Firmware</strong> <strong>Update</strong> Tool<br />
Updating the Instant Pilot<br />
Updating the Instant Pilot<br />
The Instant Pilot G4208A is updated like any other 1100/1200/1290 modules<br />
with the LAN/RS-232 <strong>Firmware</strong> <strong>Update</strong> Tool 2.7 when using firmware<br />
revsions A.06.xx/B.01.xx for the modules and B.01.xx for the Instant Pilot.<br />
If you want to downgrade an Instant Pilot to revision A.05.1x (to be<br />
compatible with modules that have A.05.09/10 or A.05.11/12 firmware<br />
installed), then follow the instructions in “<strong>Update</strong> Information for A.05.1x<br />
<strong>Firmware</strong>” on page 104.<br />
Recovering<br />
In case of power fail or LAN/RS-232 cable disconnected during the update for<br />
the Instant Pilot, the firmware may not completely updated (mian or<br />
resident). In this case the display may be dark and the LED is blinking.<br />
In this case the LAN/RS-232 <strong>Firmware</strong> <strong>Update</strong> Tool 2.7 shows the Instant Pilot<br />
either in resident mode (-R) or main mode (similar to “Main firmware update<br />
was interrupted” on page 81). Redo the firmware update again for both the<br />
main and resident (for all B.xx versions).<br />
If this happend during the downgrade to A.05.xx, the you just need to redo the<br />
main update. The firmware A.05.1x has no resident part.<br />
If the Instant Pilot is not shown, check whether the modules may be on<br />
different firmware sets (A.05.xx vs. A.06.xx/B.01.xx/B.06.xx). Then use one<br />
module with the Instant Pilot’s target firmware to establish communication.<br />
78 1100/1200/1290 Series <strong>Firmware</strong> <strong>Update</strong> Guide