<strong>Update</strong>s via Control Module (G1323A/B) 5<br />

Solving Problems<br />

Files not found on PC-card<br />

If the Control Module displays<br />

and either no files are on the PC-card or the PC-card is not recognized by the<br />

Control Module. This happens when the PC-card has been formatted using the<br />

Windows Operating system (Windows NT, Windows 2000 and Windows XP).<br />

Even if the required FAT16 is used, the PC-cards become invisible. If you run<br />

into this situation there are two possibilities for recovering of the PC-card:<br />

• Use a digital camera with PC-card slot (PCMCIA or CF-card) and use the<br />

camera’s formatting feature.<br />

• Use a tool described on the website http://www.mager.org/mkdosfs/ with<br />

the FAT16 option.<br />

Control Module Has Wrong <strong>Firmware</strong> Revision<br />

If the Control Module has the wrong firmware revision (e.g. B.03.22) and is<br />

connected to an <strong>Agilent</strong> 1100/1200 system with A.06.xx, the Control Module<br />

does not start correctly. It cannot not talk to the 1100/1200 modules due to<br />

mismatch in firmware. To overcome this situation use the “PC-Card <strong>Update</strong><br />

Mode” on page 122.<br />

1100/1200/1290 Series <strong>Firmware</strong> <strong>Update</strong> Guide 129