<strong>Update</strong>s via Control Module (G1323A/B) 5<br />
Solving Problems<br />
Files not found on PC-card<br />
If the Control Module displays<br />
and either no files are on the PC-card or the PC-card is not recognized by the<br />
Control Module. This happens when the PC-card has been formatted using the<br />
Windows Operating system (Windows NT, Windows 2000 and Windows XP).<br />
Even if the required FAT16 is used, the PC-cards become invisible. If you run<br />
into this situation there are two possibilities for recovering of the PC-card:<br />
• Use a digital camera with PC-card slot (PCMCIA or CF-card) and use the<br />
camera’s formatting feature.<br />
• Use a tool described on the website http://www.mager.org/mkdosfs/ with<br />
the FAT16 option.<br />
Control Module Has Wrong <strong>Firmware</strong> Revision<br />
If the Control Module has the wrong firmware revision (e.g. B.03.22) and is<br />
connected to an <strong>Agilent</strong> 1100/1200 system with A.06.xx, the Control Module<br />
does not start correctly. It cannot not talk to the 1100/1200 modules due to<br />
mismatch in firmware. To overcome this situation use the “PC-Card <strong>Update</strong><br />
Mode” on page 122.<br />
1100/1200/1290 Series <strong>Firmware</strong> <strong>Update</strong> Guide 129