5 <strong>Update</strong>s via Control Module (G1323A/B)<br />

General Information<br />

NOTE<br />

The G1323B Control Module received it’s last firmware revision B.04.02 in September 2005.<br />

This means that newer <strong>Agilent</strong> 1100/1200 modules are controllable with some restrictions.<br />

• G1314B Variable Wavelength Detector (*)<br />

• G1314C Variable Wavelength Detector SL (*)<br />

• G1314D Variable Wavelength Detector (*)<br />

• G1314E Variable Wavelength Detector SL+ (*)<br />

• G1316B Thermostatted Column Compartment (*)<br />

• G1316C Thermostatted Column Compartment (*)<br />

• G1329B standard autosampler SL (*)<br />

• G1367B high performance autosampler (*)<br />

• G1367C high performance autosampler SL (*)<br />

• G1367D high performance autosampler SL +(*)<br />

• G1315D Diode Array Detector (DAD) (**)<br />

• G1365D Multiple Wavelength Detector (MWD) (**)<br />

• 1290 Infinity LC System - unsupported<br />

(*) with feature set of the "A" module<br />

(**) with feature set of the "C" module (except for the high data rate)<br />

The RRLC modules G1312B Binary Pump SL and G1316B TCC SL are not supported due to<br />

it's different feature set.<br />

<strong>Update</strong> Sequence<br />

Using the Control Module, the system should be updated manually in the<br />

following order:<br />

1 <strong>Update</strong> of Control Module<br />

• G1323A has to be done via PC card and<br />

116 1100/1200/1290 Series <strong>Firmware</strong> <strong>Update</strong> Guide