6 <strong>Update</strong>s via GPIB <strong>Firmware</strong> <strong>Update</strong> Tool<br />
<strong>Firmware</strong> <strong>Update</strong> to A.06.xx<br />
Recovering<br />
In case of power fail or GPIB cable disconnected during the update to revision<br />
A.06xx, one or more modules will no longer visible in the system! It does not<br />
communicate with the modules that have A.06.xx firmware.<br />
Module has GPIB communication<br />
If the module has GPIB communication you have to<br />
1 Power off all modules.<br />
2 Add the GPIB communication to the module that is no longer visible.<br />
3 Turn this module on.<br />
4 Start the firmware update to A.06.xx for this module.<br />
5 Power off the module.<br />
6 Re-install the GPIB communication to the detector (DAD, MWD, FLD, VWD)<br />
or, if not available, to pump or autosampler (this order).<br />
7 Turn all modules on for normal operation.<br />
Module has no GPIB communication<br />
If the module has no GPIB communication (e.g. G1316A TCC) you have to<br />
1 Power off all modules except the module that hosts the GPIB<br />
communication.<br />
2 Downgrade this module to firmware A.05xx (resident and main).<br />
3 Turn the missing module on.<br />
4 Start the firmware update to A.06.xx for both modules.<br />
5 Turn all modules on for normal operation.<br />
142 1100/1200/1290 Series <strong>Firmware</strong> <strong>Update</strong> Guide