6 <strong>Update</strong>s via GPIB <strong>Firmware</strong> <strong>Update</strong> Tool<br />

<strong>Firmware</strong> <strong>Update</strong> to A.06.xx<br />

Recovering<br />

In case of power fail or GPIB cable disconnected during the update to revision<br />

A.06xx, one or more modules will no longer visible in the system! It does not<br />

communicate with the modules that have A.06.xx firmware.<br />

Module has GPIB communication<br />

If the module has GPIB communication you have to<br />

1 Power off all modules.<br />

2 Add the GPIB communication to the module that is no longer visible.<br />

3 Turn this module on.<br />

4 Start the firmware update to A.06.xx for this module.<br />

5 Power off the module.<br />

6 Re-install the GPIB communication to the detector (DAD, MWD, FLD, VWD)<br />

or, if not available, to pump or autosampler (this order).<br />

7 Turn all modules on for normal operation.<br />

Module has no GPIB communication<br />

If the module has no GPIB communication (e.g. G1316A TCC) you have to<br />

1 Power off all modules except the module that hosts the GPIB<br />

communication.<br />

2 Downgrade this module to firmware A.05xx (resident and main).<br />

3 Turn the missing module on.<br />

4 Start the firmware update to A.06.xx for both modules.<br />

5 Turn all modules on for normal operation.<br />