Appendix<br />
ISE Default Values<br />
E0 default: 0 mV<br />
Low limit of E0: - 60 mV<br />
<strong>High</strong> limit of E0: + 60 mV<br />
Slope default: - 59.2 mV/decade<br />
Thermo Scientific <strong>Orion</strong> <strong>2117HL</strong> <strong>High</strong> <strong>Level</strong> <strong>Chloride</strong> <strong>Analyzer</strong> <strong>User</strong> <strong>Guide</strong><br />
Slope range: - 62.0 mV/decade to - 45.0 mV/decade<br />
CISO: 35.5 ppm<br />
Volume of flow cell: 95 mL<br />
Concentration of standard 1 added during calibration = 2.96 %<br />
Volume of standard 1 added during calibration = 0.50 mL<br />
Concentration of standard 2 added during calibration = 2.96 %<br />
Volume of standard 2 added during calibration = 1.50 mL<br />