Calibration<br />
Error Operator Action Scrolling Marquee Main Display Notes<br />
Stability Time Out Error<br />
Error<br />
41<br />
Press to continue the calibration<br />
despite the reading instability.<br />
or<br />
Press to abort the calibration and<br />
refer to Exit 3. Repeat the calibration or<br />
refer to the troubleshooting section.<br />
Calibration Errors<br />
Error<br />
42<br />
Error<br />
107<br />
Error<br />
109<br />
Press to abort the calibration and<br />
refer to Exit 3.<br />
Verify that the correct standards were<br />
used during the calibration, the standard<br />
values were entered correctly and in the<br />
right order and the electrodes are working<br />
properly and then repeat the calibration.<br />
Press to abort the calibration and<br />
refer to Exit 3.<br />
Press to continue the calibration<br />
despite the bad slope.<br />
or<br />
Calibration<br />
Error Codes<br />
Press to abort the calibration and<br />
refer to Exit 3.<br />
READINGS WERE UNSTABLE<br />
PRESS ENTER TO CONTINUE<br />
CALIBRATION OR PRESS EXIT<br />
TO ABORT<br />
INVALID SLOPE<br />
PRESS ENTER<br />
CALIBRATION DATA TOO<br />
CLOSE TOGETHER<br />
PRESS ENTER<br />
Thermo Scientific <strong>Orion</strong> <strong>2117HL</strong> <strong>High</strong> <strong>Level</strong> <strong>Chloride</strong> <strong>Analyzer</strong> <strong>User</strong> <strong>Guide</strong><br />
At any point during a calibration, the appropriate error message will be<br />
displayed if there is a calibration related error.<br />
The following table lists common calibration error codes with their<br />
marquee display and required operator actions.<br />
CH1 BAD SLOPE<br />
PRESS ENTER<br />
E041 The required system<br />
stability was not<br />
reached and a time<br />
out error occurred.<br />
Perform electrode<br />
cleaning, conditioning<br />
and troubleshooting.<br />
E042 The calculated slope<br />
was invalid.<br />
Standard values were<br />
likely entered in the<br />
wrong sequence<br />
or wrong standard<br />
values were entered.<br />
E107 Use new standards.<br />
Perform electrode<br />
cleaning, conditioning<br />
and troubleshooting.<br />
E109 The calculated slope<br />
was outside of the<br />
recommended range.<br />
Perform electrode<br />
cleaning and<br />
conditioning, refer to<br />
the troubleshooting<br />
section and repeat the<br />
calibration.<br />
IV-11