General Information<br />

Glossary<br />

Refer to Figure I-3.<br />

Thermo Scientific <strong>Orion</strong> <strong>2117HL</strong> <strong>High</strong> <strong>Level</strong> <strong>Chloride</strong> <strong>Analyzer</strong> <strong>User</strong> <strong>Guide</strong><br />

Inlet Valve – Accepts the sample stream via 1/4 inch NPTF connector. The<br />

operator must supply the sample with a pressure between 14 and 100 psig.<br />

Inlet Filter – 60 micron stainless steel filter traps particulate matter in the<br />

sample stream.<br />

Bypass/Needle Valve – Used to redirect flow in the bypass system.<br />

Pressure Regulator – Adjusts flow on the incoming sample stream.<br />

Flow Meter – Measures the sample flow rate.<br />

Check Valve – Prevents the backflow of sample.<br />

Restrictor Tube Assembly – Used in conjunction with the pressure<br />

regulator to lower downstream pressure.<br />

Flow Cell Manifold – Directs sample flow in and out of the flow cell.<br />

Flow Cell – Contains the chloride sensing electrode, reference electrode and<br />

ATC probe.<br />

Diverter Valve – Allows the flow cell reservoir to fill during calibration by<br />

forming a closed-loop system.<br />

<strong>Chloride</strong> Electrode – Senses chloride ions in the sample stream and<br />

produce an electrical potential dependent on sample concentration.<br />

Reference Electrode – Provides a constant reference potential and<br />

completes the measurement circuit.<br />

Pressurized Reference Electrode Filling Solution Bottle – Provides<br />

constant flow of electrolyte solution through reference electrode for<br />

maximum stability.<br />

ATC Probe – Measures the sample temperature and inputs data to<br />

microprocessor for automatic temperature compensation (ATC).<br />

Calibration Port – Allows introduction of standards to the sample reservoir<br />

during calibration.<br />

Air Pump – Used to mix the sample during both measurement and<br />

calibration as well as provide air to pressurize the reference electrode filling<br />

solution bottle.<br />