Calibration<br />

Step Operator Action Scrolling Marquee Main Display Notes<br />

Calibration Abort Exit Steps – If is pressed at any point during the calibration, proceed with the following steps<br />

Exit 1 ARE YOU SURE?<br />

Exit 2 Press to abort the calibration.<br />

or<br />

Calibration<br />

Abort Steps<br />

Press to return to the last screen<br />

used in the calibration sequence prior to<br />

when was pressed. Continue to<br />

follow the calibration steps according to<br />

the marquee and main display directions.<br />

Thermo Scientific <strong>Orion</strong> <strong>2117HL</strong> <strong>High</strong> <strong>Level</strong> <strong>Chloride</strong> <strong>Analyzer</strong> <strong>User</strong> <strong>Guide</strong><br />

PRESS ENTER TO ABORT OR PRESS<br />

LAST SCREEN TO BACKUP<br />

ARE YOU SURE?<br />

PRESS ENTER TO ABORT OR PRESS<br />

LAST SCREEN TO BACKUP<br />

Exit 3 CALIBRATION ABORTED<br />

PRESS ENTER<br />

Exit 4 Press to abort the calibration. CALIBRATION ABORTED<br />

Exit 5 Pull the diverter valve out.<br />

Turn the inlet valve on.<br />

Close the calibration port on the<br />

flow cell cap.<br />

PRESS ENTER<br />

PULL DIVERTER VALVE OUT<br />

TURN INLET VALVE ON<br />

CLOSE CAL PORT<br />

PRESS ENTER<br />

Exit 6 Press . PULL DIVERTER VALVE OUT<br />

Exit 7 No operator action needed.<br />

The system will enter the measure mode<br />

and remain on hold until the hold timer<br />

expires (default hold time is 30 minutes).<br />

The system allows the operator to abort the calibration or return to the<br />

calibration when is pressed at any point during the calibration.<br />

The following table lists the calibration abort steps with their marquee<br />

display and required operator actions.<br />

TURN INLET VALVE ON<br />

CLOSE CAL PORT<br />

PRESS ENTER<br />

SUrE<br />

?<br />

SUrE<br />

?<br />

CAL<br />

Abrt<br />

CAL<br />

Abrt<br />

FLO On<br />

FLO On<br />

Actual temperature reading Actual concentration<br />

reading<br />

IV-10