General Information<br />

Principles of Operation<br />

Figure I-1<br />

<strong>2117HL</strong> Schematic<br />

Thermo Scientific <strong>Orion</strong> <strong>2117HL</strong> <strong>High</strong> <strong>Level</strong> <strong>Chloride</strong> <strong>Analyzer</strong> <strong>User</strong> <strong>Guide</strong><br />

The sample enters the Thermo Scientific <strong>Orion</strong> <strong>2117HL</strong> chloride analyzer<br />

and passes through the inlet valve, bypass/needle valve, inlet filter, pressure<br />

regulator, flow meter and into the restrictor tubing. The sample then passes<br />

through the flow cell manifold into the flow cell as air is introduced from<br />

the air pump to ensure proper mixing and fast response. The sample passes<br />

the chloride sensing electrode, reference electrode and ATC probe and then<br />

flows into an atmospheric drain via the diverter valve.<br />

Safety Drain<br />

for Overflow<br />

Power<br />

4-20mA Output<br />

<strong>Analyzer</strong><br />

1 2 3 4 5<br />

Calibration <strong>Level</strong><br />

Measure <strong>Level</strong><br />

Valve Positions<br />

REF<br />

Calibration Mode - PUSH<br />

TEMP<br />

Sample<br />

Flow Cell Block<br />

Sample/Measure Mode - PULL<br />

Drain<br />

Valve Val alve<br />

Sense<br />

Cal<br />

Reference Solution Calibration<br />

Port<br />

Air Filter<br />

Air Inlet<br />

Pressurizes Flow Cell<br />

to Initiate Siphoning<br />

Flow Cell Manifold<br />

Pressurizes Reference<br />

Filling Solution Bottle<br />

Air Pump<br />

Restrictor Tubing<br />

Main Feed<br />

Check<br />

Valve<br />

Flow Meter<br />

Regulator<br />

Valve<br />

Bypass / Needle<br />

Valve and Filter<br />

Inlet Valve<br />

K E Y<br />

Electric<br />

Fluid Path<br />

Fluid Restrictor Tubing<br />

Air Path<br />

Reference Solution Path<br />

Overflow Path<br />