Calibration<br />
Step Operator Action Scrolling Marquee Main Display Notes<br />
17 Wait 0 to 8 minutes. AWAITING DATA STABILITY Actual readings<br />
flashing<br />
18 Add standard 2 through the<br />
calibration port.<br />
Pipet 1.5 mL of the standard into the flow<br />
cell by adding three increments of the<br />
standard using the 0.5 mL pipet.<br />
ADD STD2<br />
PRESS ENTER<br />
19 Press . ADD STD2<br />
PRESS ENTER<br />
20 Wait 2 minutes. MEASURING STD2<br />
CIRCULATING<br />
WAIT 2 MINUTES<br />
Thermo Scientific <strong>Orion</strong> <strong>2117HL</strong> <strong>High</strong> <strong>Level</strong> <strong>Chloride</strong> <strong>Analyzer</strong> <strong>User</strong> <strong>Guide</strong><br />
System continuously<br />
evaluates the<br />
data until a stable<br />
condition is met.<br />
If there is a calibration<br />
error, the appropriate<br />
error message will be<br />
displayed.<br />
Add Std2 For calibration<br />
standard 2, use the<br />
29,600 ppm high level<br />
chloride standard<br />
(Cat. No. 211740HL).<br />
Add Std2<br />
Actual readings<br />
flashing<br />
21 Wait 0 to 8 minutes. AWAITING DATA STABILITY Actual readings<br />
flashing<br />
System circulates<br />
sample. Minutes will<br />
be counted down on<br />
the marquee.<br />
System continuously<br />
evaluates the<br />
data until a stable<br />
condition is met.<br />
If there is a calibration<br />
error, the appropriate<br />
error message will be<br />
displayed.<br />
22 No action necessary. System calculates<br />
new Eo and slope and<br />
checks for errors.<br />
If there is a calibration<br />
error, the appropriate<br />
error message will be<br />
displayed.<br />
IV-8