<strong>Analyzer</strong> Operation<br />

TEST<br />

CAL<br />

DKA<br />

Thermo Scientific <strong>Orion</strong> <strong>2117HL</strong> <strong>High</strong> <strong>Level</strong> <strong>Chloride</strong> <strong>Analyzer</strong> <strong>User</strong> <strong>Guide</strong><br />

Test relays and analog output lines (DIAGNOSTICS will appear in the<br />

mode indicator):<br />

• Method to activate/deactivate relays and outputs to be tested<br />

• Verify the accuracy of the analog outputs when used with an external<br />

loop calibrator<br />

• Provides the values and settings for the mA output and relays<br />

• mA output<br />

• 4-20 or 0-20<br />

• The sensor (SOUt) and temperature (tOUt) low and high values<br />

• Relay status<br />

• Set RLY1, RLY2 and RLY3 status to OFF or On<br />

Set calibration frequency in hours:<br />

• <strong>High</strong> limit is 19999 hours<br />

• Low limit is 00000 hours<br />

• Default setting – 720 hours<br />

Set values for customized Double Known Addition (DKA):<br />

• Programmable for volume (mL) of flow cell, concentration (ppm) and<br />

volume (mL) of standard 1, concentration (ppm) and volume (mL) of<br />

standard 2 and volume (mL) of additions<br />

• Default flow cell volume: 95.0 mL<br />

• Default concentration (Std1): 2.96 %*<br />

• Default volume addition (Std1): 0.5 mL<br />

• Default concentration (Std2): 2.96 %*<br />

• Default volume addition (Std2): 1.5 mL**<br />

* Use the 29,600 ppm high level chloride standard (Cat. No.<br />

211740HL) as both standard 1 (Std1) and standard 2 (Std2)<br />

during the calibration of the <strong>2117HL</strong> analyzer. The analyzer will<br />

express the standard value in percent (parts per hundred) in the<br />

DKA setup menu and during a DKA calibration.<br />

** The standard 2 volume addition is made by adding three<br />

increments of standard 2 to the flow cell using the 0.5 mL pipet.<br />