Calibration<br />

Step Operator Action Scrolling Marquee Main Display Notes<br />

10 Perform laboratory analysis on QC sample<br />

and return with results.<br />

Thermo Scientific <strong>Orion</strong> <strong>2117HL</strong> <strong>High</strong> <strong>Level</strong> <strong>Chloride</strong> <strong>Analyzer</strong> <strong>User</strong> <strong>Guide</strong><br />

Actual temperature reading Actual concentration<br />

reading<br />

System will return to<br />

measure mode and<br />

operate as normal.<br />

CAL and MEASURE<br />

appear as mode.<br />

11 Press to resume the calibration. HOLD and CAL appear<br />

as mode.<br />

12 If the calibration password is active, enter<br />

the calibration password as described in<br />

step 3.<br />

Press to accept the password.<br />

13 CH1 or CH2<br />

14<br />

15<br />

Press / to select the<br />

channel that will be calibrated and press<br />

to begin the calibration.<br />

Press / to select OFFL<br />

and press to accept the setting.<br />

16 Enter the concentration value of the QC<br />

sample obtained from laboratory analysis:<br />

Press to move the decimal point<br />

and set the value as ppm or ppb.<br />

Press / to set the first digit.<br />

Press to move to the next digit.<br />

Press / to set the values of<br />

the remaining digits and press to<br />

move through the remaining digits.<br />

Press to save the value.<br />

ENTER PASSWORD PASS<br />

(flashing)<br />

CH1 or CH2<br />

(flashing)<br />

OFFL<br />

(flashing)<br />

LAB ANALYSIS READY<br />

INPUT CONC VALUE FROM LAB<br />

PRESS ENTER TO CONTINUE OR<br />

PRESS EXIT TO ABORT<br />

17 NEW CALIBRATION Eo<br />

PRESS ENTER TO ACCEPT OR<br />

PRESS EXIT TO ABORT<br />

# # # #<br />

CAL<br />

CHnL<br />

CAL<br />

CHnL<br />

CAL<br />

tYPE<br />

COnC<br />

# # # . #<br />

(change flashing digit)<br />

Displayed for two<br />

channel analyzer<br />

Action required for<br />

two channel analyzer<br />

Actual Eo value The system<br />

will display the<br />

recalculated the E 0<br />

18 Press . The system will return<br />

to the measure mode.<br />

Only MEASURE<br />

should appear<br />

as mode.<br />

IV-16