Orion 2117HL High Level Chloride Analyzer User Guide
Orion 2117HL High Level Chloride Analyzer User Guide
Orion 2117HL High Level Chloride Analyzer User Guide
You also want an ePaper? Increase the reach of your titles
YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.
Appendix<br />
Outputs<br />
Thermo Scientific <strong>Orion</strong> <strong>2117HL</strong> <strong>High</strong> <strong>Level</strong> <strong>Chloride</strong> <strong>Analyzer</strong> <strong>User</strong> <strong>Guide</strong><br />
Number of Analog Outputs Two, one dedicated to chloride, one dedicated to<br />
temperature and shared ground<br />
Analog Outputs Galvanically isolated<br />
Output Selections 0 to 20 mA or 4 to 20 mA<br />
Programmable Range Yes<br />
Channel Assignments <strong>Chloride</strong> and temperature<br />
Relative Accuracy ± (0.05 mA + 0.5%)<br />
Maximum Load 500 ohm or 10 V<br />
Log and Linear Output Options Yes, operator selectable<br />
Alarm Outputs<br />
Number of Relay Outputs 3<br />
Maximum Relay Load 250 VAC/5 A, 30 VDC/5 A, resistive load only<br />
Minimum Value Alarm Yes<br />
Maximum Value Alarm Yes<br />
Error Alarm Yes<br />
Calibration/offline Alarm Yes<br />
Programmable Min. And Max. Values Yes<br />
<strong>Analyzer</strong> Features<br />
Non-volatile Memory Yes<br />
Battery Backup (Replaceable) Yes<br />
Data Logging Yes<br />
Keypad 8 electromechanical keys with graphic overlay<br />
Software Features<br />
Self-test and Diagnostics Yes<br />
Real Time Clock Yes<br />
Password Protection Yes<br />
Reset Function Yes<br />
Programmable Alarms <strong>High</strong>, low, error, calibration/offline<br />
Measurement, Active Signal noise, temperature<br />
Calibration Sensor response, drift, E 0, slope, temperature<br />
A-5