Appendix<br />

Outputs<br />

Thermo Scientific <strong>Orion</strong> <strong>2117HL</strong> <strong>High</strong> <strong>Level</strong> <strong>Chloride</strong> <strong>Analyzer</strong> <strong>User</strong> <strong>Guide</strong><br />

Number of Analog Outputs Two, one dedicated to chloride, one dedicated to<br />

temperature and shared ground<br />

Analog Outputs Galvanically isolated<br />

Output Selections 0 to 20 mA or 4 to 20 mA<br />

Programmable Range Yes<br />

Channel Assignments <strong>Chloride</strong> and temperature<br />

Relative Accuracy ± (0.05 mA + 0.5%)<br />

Maximum Load 500 ohm or 10 V<br />

Log and Linear Output Options Yes, operator selectable<br />

Alarm Outputs<br />

Number of Relay Outputs 3<br />

Maximum Relay Load 250 VAC/5 A, 30 VDC/5 A, resistive load only<br />

Minimum Value Alarm Yes<br />

Maximum Value Alarm Yes<br />

Error Alarm Yes<br />

Calibration/offline Alarm Yes<br />

Programmable Min. And Max. Values Yes<br />

<strong>Analyzer</strong> Features<br />

Non-volatile Memory Yes<br />

Battery Backup (Replaceable) Yes<br />

Data Logging Yes<br />

Keypad 8 electromechanical keys with graphic overlay<br />

Software Features<br />

Self-test and Diagnostics Yes<br />

Real Time Clock Yes<br />

Password Protection Yes<br />

Reset Function Yes<br />

Programmable Alarms <strong>High</strong>, low, error, calibration/offline<br />

Measurement, Active Signal noise, temperature<br />

Calibration Sensor response, drift, E 0, slope, temperature<br />