Calibration<br />

Step Operator Action Scrolling Marquee Main Display Notes<br />

9 Press . CHECK CAL PORT CLOSED<br />

10 When the liquid level in the flow cell<br />

begins to drop, turn the inlet valve off.<br />

Open the calibration port on the flow<br />

cell cap.<br />

PUSH DIVERTER VALVE IN<br />

PRESS ENTER<br />

Thermo Scientific <strong>Orion</strong> <strong>2117HL</strong> <strong>High</strong> <strong>Level</strong> <strong>Chloride</strong> <strong>Analyzer</strong> <strong>User</strong> <strong>Guide</strong><br />

WHEN LIQUID LEVEL BEGINS TO DROP<br />

TURN INLET VALVE OFF<br />

OPEN CAL PORT<br />

PRESS ENTER<br />

11 Press . WHEN LIQUID LEVEL BEGINS TO DROP<br />

TURN INLET VALVE OFF<br />

OPEN CAL PORT<br />

PRESS ENTER<br />

12 Wait 2 minutes. STARTING CALIBRATION<br />

MEASURING BACKGROUND<br />

WAIT 2 MINUTES<br />

FILL<br />

FCEL<br />

FLO OFF<br />

FLO OFF<br />

Actual readings<br />

flashing<br />

13 Wait 0 to 8 minutes. AWAITING DATA STABILITY Actual readings<br />

flashing<br />

14 Add standard 1 through the<br />

calibration port.<br />

Pipet 0.5 mL of the standard into the flow<br />

cell using the 0.5 mL pipet.<br />

ADD STD1<br />

PRESS ENTER<br />

15 Press . ADD STD1<br />

PRESS ENTER<br />

16 Wait 2 minutes. MEASURING STD1<br />

CIRCULATING<br />

WAIT 2 MINUTES<br />

System circulates<br />

sample. Minutes will<br />

be counted down on<br />

the marquee.<br />

System continuously<br />

evaluates the<br />

data until a stable<br />

condition is met.<br />

If there is a calibration<br />

error, the appropriate<br />

error message will be<br />

displayed.<br />

Add Std1 For calibration<br />

standard 1, use the<br />

29,600 ppm high level<br />

chloride standard<br />

(Cat. No. 211740HL).<br />

Add Std1<br />