Appendix<br />

Specifications<br />

<strong>Chloride</strong> Measurement<br />

Thermo Scientific <strong>Orion</strong> <strong>2117HL</strong> <strong>High</strong> <strong>Level</strong> <strong>Chloride</strong> <strong>Analyzer</strong> <strong>User</strong> <strong>Guide</strong><br />

Range 75 ppm to 1000 ppm<br />

Accuracy (with DKA calibration) ± 10%<br />

Resolution 2, 3 or 4 significant digits<br />

Response Time 50% of final reading within 2 minutes of injecting a<br />

standard solution<br />

Units Displayed ppb, ppm (auto ranging)<br />

Temperature Compensation Automatic<br />

mV Measurement<br />

Range ± 1999.9 mV<br />

Resolution 0.1 mV<br />

Relative Accuracy ± (0.5 mV + 0.1%)<br />

Temperature Measurement<br />

Range -10 to 120 °C<br />

Resolution 0.1 °C<br />

Relative Accuracy ± 0.5 °C<br />

Failure Detection Manual compensation<br />

ATC Probe 30 K thermistor<br />

LED Status Indicator<br />

Green OK<br />

Yellow Warning<br />

Red Failure<br />

A-3