Troubleshooting<br />

Troubleshooting<br />

Matrix<br />

Thermo Scientific <strong>Orion</strong> <strong>2117HL</strong> <strong>High</strong> <strong>Level</strong> <strong>Chloride</strong> <strong>Analyzer</strong> <strong>User</strong> <strong>Guide</strong><br />

Malfunction Possible Cause Remedy<br />

Noisy Blocked junction on<br />

reference electrode<br />

<strong>Chloride</strong> or reference<br />

electrode failure<br />

Make sure electrolyte is flowing properly and<br />

the bottle is pressurized by the air pump.<br />

Check the cable connections and location of<br />

wiring to terminal strip. Refer to Chapter II.<br />

Temperature probe failure Disconnect ATC. Temperature should read<br />

25 °C (E101). Replace if necessary.<br />

Air pump failure Check the air flow for a steady stream of<br />

bubbles in the flow cell and replace the air<br />

pump if necessary.<br />

Calibration standard<br />

carryover<br />

Disconnect the air pump intake tube. Rinse<br />

the flow cell and electrodes with deionized<br />

water by lifting the flow cell cap. Reconnect<br />

the air pump.<br />

Faulty electrode cables Replace electrode cables.<br />

Excessive Drift Sample concentration<br />

varying<br />

Blocked junction on<br />

reference electrode<br />

<strong>Chloride</strong> or reference<br />

electrode failure<br />

Reference electrode<br />

filling solution empty or<br />

incorrect solution used<br />

Check sample status. Perform a span check<br />

and if span check passes, the analyzer is okay.<br />

Make sure electrolyte is flowing properly and<br />

the bottle is pressurized by the air pump.<br />

Check the cable connections and location of<br />

wiring to terminal strip. Refer to Chapter II.<br />

Use Cat. No. 100057F4 reference electrode<br />

filling solution.<br />

Temperature probe failure Disconnect ATC. Temperature should read<br />

25 °C (E101). Replace if necessary.<br />

Low Flow Rate Sample pressure below<br />

8 psi<br />

Pressure regulator set<br />

too low<br />

Check the sample pressure. If it is less than<br />

8 psi, increase the pressure. Refer to<br />

Chapter II.<br />

Increase the pressure by pulling on the<br />

red locking ring and turning the black knob<br />

clockwise. Refer to Chapter II.<br />

Inlet filter clogged Replace or clean the filter.<br />

Restrictor tubing crimped<br />

or clogged<br />

Replace the restrictor tubing. Refer to<br />

Chapter II.<br />

VI-11