30.11.2012 Views

Orion 2117HL High Level Chloride Analyzer User Guide

Orion 2117HL High Level Chloride Analyzer User Guide

Orion 2117HL High Level Chloride Analyzer User Guide

SHOW MORE
SHOW LESS

Create successful ePaper yourself

Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.

<strong>Analyzer</strong> Operation<br />

RSET<br />

DISP<br />

Thermo Scientific <strong>Orion</strong> <strong>2117HL</strong> <strong>High</strong> <strong>Level</strong> <strong>Chloride</strong> <strong>Analyzer</strong> <strong>User</strong> <strong>Guide</strong><br />

Reset the analyzer to factory defaults for setup parameters:<br />

• Use to troubleshoot the system (a hard reset can be performed if the<br />

keypad and software are not responding, refer to Chapter VI, Resetting<br />

the <strong>Analyzer</strong>)<br />

Warning: Resetting the analyzer will lose all stored information including<br />

relay, logs and calibration settings.<br />

Set the automatic lighting options for the backlit display:<br />

• AUtO – Brightness will change in response to ambient light source<br />

• On – Backlit display is always on<br />

• OFF – Backlit display is always off<br />

• Default display – AUtO<br />

III-6

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!