30.11.2012 Views

Orion 2117HL High Level Chloride Analyzer User Guide

Orion 2117HL High Level Chloride Analyzer User Guide

Orion 2117HL High Level Chloride Analyzer User Guide

SHOW MORE
SHOW LESS

Create successful ePaper yourself

Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.

<strong>Analyzer</strong> Maintenance<br />

Replacement of<br />

Restrictor Tubing<br />

Figure V-2<br />

Restrictor Tubing<br />

Thermo Scientific <strong>Orion</strong> <strong>2117HL</strong> <strong>High</strong> <strong>Level</strong> <strong>Chloride</strong> <strong>Analyzer</strong> <strong>User</strong> <strong>Guide</strong><br />

Use the restrictor tube assembly (Cat. No. 2100RT). Refer to Figure V-2<br />

for the restrictor tubing connections.<br />

1. Close the inlet valve to stop sample flow through the analyzer.<br />

2. Loosen the restrictor tube fitting connection from the top of the<br />

flow meter.<br />

3. Loosen the restrictor tube fitting from the left side of the flow<br />

cell block.<br />

4. Pull out the restrictor tube assembly at each end.<br />

5. Unscrew the clip behind flow cell that holds the restrictor tube to<br />

the panel.<br />

6. Replace with a new restrictor tubing assembly.<br />

7. Be sure all of the fittings are secure, to prevent leaks.<br />

8. Open the inlet valve to start sample flow through the analyzer.<br />

restrictor<br />

adapter<br />

fitting<br />

restrictor<br />

tube assembly<br />

flow<br />

meter<br />

clamp<br />

O-ring<br />

to check valve on<br />

flow cell manifold<br />

V-5

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!