30.11.2012 Views

Orion 2117HL High Level Chloride Analyzer User Guide

Orion 2117HL High Level Chloride Analyzer User Guide

Orion 2117HL High Level Chloride Analyzer User Guide

SHOW MORE
SHOW LESS

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

Calibration<br />

Before Performing a<br />

DKA Calibration<br />

Recommendations<br />

Figure IV-4<br />

Pipet Dispensing<br />

Thermo Scientific <strong>Orion</strong> <strong>2117HL</strong> <strong>High</strong> <strong>Level</strong> <strong>Chloride</strong> <strong>Analyzer</strong> <strong>User</strong> <strong>Guide</strong><br />

Maximum system accuracy is ensured through a fast, easy and accurate<br />

calibration performed in the expected sample range. This calibration<br />

procedure uses equipment supplied in the calibration kit<br />

(Cat. No. <strong>2117HL</strong>CAL).<br />

• Wait at least one hour to perform a calibration after cleaning<br />

the electrode.<br />

• Check that the present sample concentration is equal or less than<br />

standard 1. If not, calibration can still be performed with a loss in<br />

accuracy as long as the background concentration is not significantly<br />

larger than the concentration of standard 1. The best results are<br />

achieved when standard 1 is twice the level of the present sample<br />

concentration.<br />

• Proper pipet technique is crucial to the calibration. If the operator<br />

is not familiar with the pipet equipment or technique, refer to the<br />

Appendix, Pipet Operation section.<br />

• Do not touch the pipet tip with bare fingers. Risk of contamination is<br />

possible. Use powder free gloves to prevent contamination.<br />

• When using a pipet to inject standards during calibration, use a fresh<br />

pipet tip for each standard and each injection.<br />

• When a fresh tip is installed, rinse the tip with the same concentration<br />

standard as the intended injection.<br />

• Fill the tip and dispense the standard 1 to 3 times before injecting the<br />

standard into the flow cell. Be sure to discard the standard in the<br />

drain between the rinses.<br />

• Wipe any excess liquid from the outside of the tip with out touching<br />

the opening.<br />

• Insert the pipet tip into calibration port on flow cell cap.<br />

• Be sure to hold the pipet completely vertical while dispensing. Do not<br />

inject the standard on the side of the flow cell, as a film may be retained<br />

on the inside of the wall, resulting in errors in precision of calibration.<br />

Refer to Figure IV-4.<br />

IV-5

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!