<strong>Analyzer</strong> Maintenance<br />

Monthly Maintenance<br />

Calibration<br />

Replacement of Reference<br />

Electrode Filling Solution<br />

Thermo Scientific <strong>Orion</strong> <strong>2117HL</strong> <strong>High</strong> <strong>Level</strong> <strong>Chloride</strong> <strong>Analyzer</strong> <strong>User</strong> <strong>Guide</strong><br />

Calibration frequency is operator dependent for the most accurate and<br />

precise measurements.<br />

• Refer to Chapter IV, Performing a DKA Calibration.<br />

• Calibration intervals can be programmed using the setup mode. This<br />

will prompt the operator to recalibrate after a specific amount of time<br />

has passed. Refer to Chapter III, Use of the Setup Mode.<br />

Refer to Chapter II, Installation of a New Reference Electrode.<br />

Reference electrode filling solution is meant to be a several month supply.<br />

Therefore, if the filling solution bottle is less than one quarter full, replace<br />

the solution. This should be checked each month to ensure that the<br />

electrode does not run dry.<br />

Danger: Turn off the air pump prior to removing the reference electrode<br />

filling solution bottle. If the air pump is left on, it will spatter filling<br />

solution as the bottle is removed.<br />

1. Turn off the air pump. Refer to Chapter III Shutdown and<br />

Start-Up Procedure.<br />

2. Remove the hollow push pin on the discharge tube of the air pump<br />

from the base of the filling solution bottle.<br />

3. Remove the filling solution bottle from the clip. Unscrew the cap and<br />

discard the bottle.<br />

4. Remove the cap and seal from a new reference electrode filling solution<br />

bottle. Check that the rubber gasket is properly aligned, and then<br />

connect the cap and tubing assembly to the bottle. The 1/8” tubing<br />

should extend into the bottle.<br />

5. Invert the filling solution bottle and snap it into the clip.<br />

6. Puncture the base of the filling solution bottle with the hollow push pin<br />

on the discharge tube of the air pump. Push the pin down until its<br />

PVC base abuts the bottle. Mark and date the level of filling solution<br />

in the reservoir. The electrolyte solution will begin to flow into the<br />

reference electrode.<br />

V-2