<strong>Analyzer</strong> Maintenance<br />

Replacement of<br />

Sample Inlet Filter<br />

Thermo Scientific <strong>Orion</strong> <strong>2117HL</strong> <strong>High</strong> <strong>Level</strong> <strong>Chloride</strong> <strong>Analyzer</strong> <strong>User</strong> <strong>Guide</strong><br />

1. If the sample flow can be adjusted with the pressure regulator to<br />

50 mL/minute or more, skip this procedure.<br />

2. Turn off sample flow by turning the inlet valve to the right.<br />

3. Remove the filter hex cap using a 1 inch open-ended wrench.<br />

4. Separate the filter cap with bypass/needle valve assembly from the filter<br />

body, retaining the ring.<br />

5. Remove the filter element and soft gasket (if necessary).<br />

6. Replace the gasket and press a new sintered element into the filter body.<br />

7. Replace the hex cap and bypass/needle valve assembly and reposition<br />

the bypass drain line into the drain assembly.<br />

8. Retighten the hex cap to approximately 75 pound-inch in torque.<br />

9. Turn on the sample flow by turning the inlet valve left.<br />

10. Check for leaks.<br />

V-3